半(ban)自動(dong)頂空(kong)進(jin)樣(yang)器(qi)是(shi)壹(yi)種氣(qi)相色譜分析(xi)的(de)樣(yang)品前(qian)處(chu)理(li)裝(zhuang)置,與(yu)其他(ta)的樣(yang)品預處理(li),如:熱解(jie)吸等(deng)方(fang)法(fa)比(bi)較(jiao),更(geng)簡(jian)單(dan)實(shi)用(yong),工(gong)作(zuo)效率(lv)更(geng)高,色譜分析(xi)結果重現性(xing)好(hao)等(deng)優點(dian),是(shi)氣(qi)相色譜痕(hen)量分析(xi)的(de)優(you)選方(fang)法(fa)之壹(yi)。進(jin)樣(yang)針(zhen)頭(tou)可(ke)自由更換,方(fang)便(bian)連(lian)接國內(nei)外各種(zhong)型(xing)號(hao)的(de)氣(qi)相色譜儀(yi)。是(shi)將待(dai)測(ce)樣(yang)品放(fang)置於(yu)壹(yi)密閉(bi)容器(qi)中(zhong),通(tong)過加熱升(sheng)溫使待(dai)測(ce)樣(yang)品中(zhong)揮(hui)發(fa)性(xing)組(zu)分從樣(yang)品中(zhong)揮(hui)發(fa)出(chu)來,在氣(qi)—液(ye)(或(huo)氣(qi)—固)兩(liang)相中(zhong)達到平(ping)衡(heng),直(zhi)接抽取(qu)上(shang)部(bu)氣(qi)體進(jin)行(xing)色譜分析(xi),從(cong)而檢測(ce)出(chu)樣品中(zhong)可(ke)揮(hui)發(fa)性(xing)的組(zu)分和含(han)量。
1、儀器(qi)安裝(zhuang)
開箱後,將儀器擺放(fang)到(dao)GC的旁(pang)邊(bian)(靠近(jin)汽(qi)化(hua)室(shi)的壹(yi)邊(bian)),把(ba)供給GC載(zai)氣(qi)的氣(qi)路加三通(tong)分出(chu)壹(yi)路供給頂(ding)空(kong)進(jin)樣(yang)器(qi)(接載氣(qi)入(ru)口(kou)),堵住載氣(qi)入(ru)口(kou);關閉(bi)色譜儀(yi)載(zai)氣(qi)壓力閥(fa);把(ba)供給GC空(kong)氣(qi)的氣(qi)路加三通(tong)分出(chu)壹(yi)路供給頂(ding)空(kong)進(jin)樣(yang)器(qi)(接空氣(qi)入(ru)口(kou))連接儀器電(dian)源線(220VAC)及(ji)儀(yi)器同步(bu)啟(qi)動(dong)信號(hao)線,紅黃(huang)為輸出(chu),藍綠為(wei)輸入(ru)。
2、儀器操作(zuo)
①啟(qi)動(dong)氣(qi)相色譜儀(yi)到(dao)正常的工作(zuo)狀(zhuang)態;
②啟(qi)動(dong)頂空(kong)進(jin)樣(yang)器(qi)進(jin)入(ru)工作(zuo)狀(zhuang)態;
③把(ba)頂空(kong)進(jin)樣(yang)器(qi)的(de)樣品傳送(song)管線的(de)針(zhen)頭(tou)插入(ru)GC的進(jin)樣(yang)口(kou)中(zhong),記下(xia)GC的(de)柱前(qian)壓(ya)力,然後適(shi)當調(tiao)低GC的(de)載(zai)氣(qi)後,再調(tiao)節頂(ding)空(kong)進(jin)樣(yang)器(qi)的(de)載氣(qi)壓力至(zhi)GC所需(xu)載氣(qi)壓力,然後再關掉(diao)GC的(de)載氣(qi);
④調節加壓氣(qi)的穩(wen)壓(ya)閥(fa)(左(zuo)側),使其壓力(li)大於(yu)頂(ding)空(kong)進(jin)樣(yang)器(qi)的(de)載氣(qi)壓力(右(you)側),通(tong)常(chang)加壓壓力比(bi)載(zai)氣(qi)壓力高0.02Mpa左(zuo)右(you);若(ruo)使用GC的載(zai)氣(qi)為頂空進(jin)樣(yang)器(qi)提(ti)供進(jin)樣(yang)載(zai)氣(qi),請與(yu)廠(chang)家聯(lian)系;
⑤實驗(yan)結束後關電(dian)關氣(qi)。